225W (10CFM forced air) 140W (natural convection) Ratings are inclusive of aux output power
Turn-on Delay:
< 1 second @ 85VAC
Output Voltage Tolerance:
± 2.0% (worst case, set point + load/line regulation)
Load Regulation:
± 0.5% (measured at connector)
Line Regulation:
± 0.2% (measured at connector)
Ripple:
1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:
<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:
<5% (full load)
Hold-up Time:
10ms min. (full load)
Aux Output Voltage Regulation:
12V ± 2V (10V min. / 14V max.), for 0A ≤ IAux ≤1.2A
Aux regulation guaranteed when main output power ≥ 15% of 225W rating (~34W).
For main output power < 15%, aux output voltage may drop to < 50% of nominal rating, depending on aux output load. Please check with your system. For main output = 12V, aux output voltage is same as main output.
Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:
EN61000-3-2, Class A
Voltage Fluctuations/Flicker:
EN61000-3-3
Electrostatic Discharge (ESD) Immunity:
EN61000-4-2, 10KV contact discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:
EN61000-4-3, 20V/m 80-2700MHz, 80% 1KHz AM, 80% 5Hz AM
EFT/Burst Immunity:
EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:
EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:
EN61000-4-6, 3VRMS, (6VRMS in ISM/amateur bands), 80% 1KHz AM
Power Frequency Magnetic Field Immunity:
EN61000-4-8, 30A/m
Voltage Dip Immunity:
EN61000-4-11
Mechanical Formats
GTM962250P:
Open-frame No suffix: 2" x 4" style, 101.6*50.8*30.0mm, 170g -C suffix: 3" x 5" style, 127.0*64.8*30.0mm, 180g -D suffix: 4.22" x 7" style, 177.8*107.2*30.0mm, 190g
GTM962251P:
L-frame: 103.5*62.3*40.0mm, 230g
GTM962252P:
Cage: 103.5*62.3*40.0mm, 250g
Note 1:
The open-frame format meets the above EMC specifications as a standalone unit. However, an L-frame or cage version may help to further improve EMI performance, or to help address other system level electrical/mechanical requirements.
Input/Output Connectors
Earth Terminal (J0):
Connector on PCB: Kang Yang PCH250 (0.250" x 0.031" Faston) Recommended mate: TE Connectivity/AMP 2-520184-2 Note: Leave earth terminal disconnected for Class II input
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
Attention Mark / Refer to operating manual for operating and service instructions
No Logo Applicable
CB for IEC 60601-1:2005 (Third Edition) + CORR. 1 (2006) + CORR. 2
(2007) + AM1 (2012) or IEC 60601-1 (2012 reprint) (Ed 3.1) up to 48V only
No Logo Applicable
CB for IEC 60950-1:2005 (Second Edition) + Am 1:2009 + Am 2:2013
No Logo Applicable
CB for IEC 62368-1:2014 (Second Edition)
CCC GB17625.1-2012, GB4943.1-2011, GB/T9254-2008
CE Certification
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL
62368-1:2014 Ed.2]
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA
C22.2#62368-1:2014 Ed.2]
CHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Certification # TC RU AR46.B.75423 Custom Union of Russia, Belarus and Kazakhstan
Medical Elec. Equip.- Part 1-11: Gen. Req. For Basic Safety & Essential Perf.- Collateral Standard - Req.
For Medical Elec. Equip. & Medical Elec. Systems Used In The Home Healthcare Environment [IEC 60601-
1-11:2015 Ed.2]
Indoor Use Only - Mark is on the label or Molded in the case
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
Japan: Voluntary Control Council for Interference (VCCI)
WEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm)
Mark is on the label or Molded in the case
Popular Products (Stock items)
SKU
Volts Out
Current Out (Amps)
Connector
RFQ
Sample Request
1852843
24.0
9.375
1686487
54.0
4.16
1686493
12.0
18.75
1686486
12.0
18.75
1686490
54.0
4.16
Hva er en open frame eller intern AC/DC strømforsyning ?
Open frame AC/DC strømforsyninger er også kjent som
AC/DC convertere og er en type intern strømforsyning.
Open Frame strømforsyningen er bygget inn i produktet
som skal strømforsynes og benyttes til å omdanne AC strøm
til DC spenning. Hva er et IEC 62368-1:2018 power supply `?
IEC 62368-1:2018er benyttes til sikkethet for elektriske og elektroniske apparater angående audio, video , informasjon og kommunikasjonsteknologi, business og kontormaskiner med en gradert spenning som ikke overtiger 600V. Dette er en produktsikkerhet standard som klassifiserer energi kilder, foreskrevne beskyttelser mot disse energi- kilder og gir veiledning ved bruk av - og krav til disse beskyttelser. De foreskrevne beskyttelseskrav er ment å redusere sannsynlighet for smerte, injurier og i tilfelle brann - eiendomsskader. Et IEC 62368 godkjent AC/DC power supply , er et adapter som er testet og sertifisert i henhold til gitte standard.
Erstatter et IEC 62368-1 godkjent power supply et IEC 60950-1 godkjent power supply ? 20 desember 2020 er sikkerhets- standardene IEC 60950-1 og IEC 60065 historie og erstattes av IEC 62368-1 vedrørende tekniske- og audio video produkter. I nord-Amerika vil ETL og UL60950-1 godkjenninger fortsatt gjelde og vil antageligvis bli benyttet inntil videre. Hva er Level VI effektivitet ?
Nummerene I til VI er en internasjonal effektmerkeprotokoll
som tillater produsenter å designe effektivitetsnivået av et
produkt . Dette er basert på målinger av energieffektiviteten
og forbruk ved no load av eksterne AC/DC adaptere eller
power supply som testes ved ulike lastnivå. No load
forbruket refererer til forbruk som et eksternt power supply
trekker selv om produktet som skal strømforsynes er slått av
eller koplet ut.
Er Level VI et krav ?
Level VI effektivitet er obligatorisk i mange land og regioner
for spesielle produkter eller applikasjoner.
Hva er ECO Design Directive 2009/125/EC ?
ECO-Design Direktivet er lovbelagt i Europa i de siste
energikrav for bevaring av AC/DC adaptere og eksterne
power. Hvordan virker et AC/DC adapter ?
AC/DC adapteret eller eksternt power supply består av et
power supply som er beskyttet i et plasthus tilsvarende en
AC plugg, derav navn som plug type, wall plug in,plug in eller
plugtop adapter.
Adapteret blir montert i vegguttaket og AC spenning blir så
omdannet av AC/DC adapteret til DC spenning som er
koplet til apparatet via en utgangskabel.